Table. 2.

Table. 2.

Comparison analysis of Agility with DS2 and manual methods

Serial no. of instrumentNo. of testNo. of concordant casesNo. discordant casesOverall concordance rate (%)
1334534530100.0
134347346199.7
1372322320100.0
Total1,0321,031199.99