Report On Proficiency Testing

J Lab Med Qual Assur 2018; 40(3): 128-135

Published online September 30, 2018

https://doi.org/10.15263/jlmqa.2018.40.3.128

Copyright © Korean Association of External Quality Assessment Service.

Annual Report on the External Quality Assessment Scheme for Urinalysis and Fecal Occult Blood Testing in Korea (2017)

Chang-Ho Jeon, A-Jin Lee, Sang-Gyung Kim, Hun-Seok Suh, Young-Cheol Bae

Department of Laboratory Medicine, School of Medicine, Catholic University of Daegu, Daegu, Korea

Correspondence to:Chang-Ho Jeon Department of Laboratory Medicine, Daegu Catholic University Medical Center, School of Medicine, Catholic University of Daegu, 33 Duryugongwon-ro 17-gil, Nam-gu, Daegu 42472, Korea el: +82-53-650-4144 Fax: +82-53-653-8672 E-mail: chjeon@cu.ac.kr
교신저자: 전창호 우)42472 대구시 남구 두류공원로 17길 33, 대구가톨릭대학교병원 진단검사의학과 Tel: 053)650-4144, Fax: 053)653-8672, E-mail: chjeon@cu.ac.kr

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

In 2017, external quality assessment trials for urinalysis and fecal occult blood (FOB) were performed with 1,544 participants. Urine chemistry tests were performed three times and urine sediment and FOB tests were evaluated 2 times. Urine chemistry tests consisted of pH, protein, glucose, ketone, bilirubin, blood, urobilinogen, nitrite, leukocyte, and specific gravity analyses. The results of the urine chemistry and specific gravity tests showed accuracy rates >95%. The accuracy rate of urine sediments was low, especially for fat droplets and atypical uric acid crystals. In the FOB quality test, all reagents showed accuracy rates >82%, which suggested the persistent improvement of false-positive reactions. In the FOB quantitative test, discrepant results depending on the instrument used were observed. To compensate for the result differences caused by the amounts of stool samples, the results should be reported using another unit (µg/g of stool).

Keywords: Quality control, Urinalysis, Occult blood

요경검학분과의 2017년도 외부정도관리사업은 새로운 전산입력시스템이 안정화되어 결과분석 및 보고가 제대로 이루어졌다. 요화학검사는 3차 시행하였고 총 12개의 관리물질을 사용하였다. 요침사검사는 2회 시행하였으며 총 8개의 사진을 게시하여 판독하도록 하였다. 대변잠혈검사(faecal occult blood, FOB) 또한 2회 시행하였으며 총 6개의 관리물질을 사용하였다. 요화학검사 및 FOB 신빙도조사에는 국내에서 제조된 관리물질을 사용하였다.

2017년 신빙도조사 사업결과에서 요화학검사는 98% 이상의 정답률을 나타내었고 기기회사별로도 정답률에 큰 차이가 없었다. 그러나 Beckmann (Beckman Coulter Inc., Brea, CA, USA) 장비로 판독한 백혈구검사에서 낮은 정답률(75%)을 보였다. 요침사검사에서는 67.8%-86.7%의 정답률을 나타내어 개선의 여지가 필요하였다. FOB에서는 정답률이 90% 이상으로 위양성률이 높았던 2015년 이후 정답률의 개선이 지속되었다[1].

1. 대상

요경검학분과에서는 대한임상검사정도관리협회에 등록된 기관을 대상으로 2017년도 4월, 6월, 8월 3차에 걸쳐 정도관리물질을 발송하였다. 발송기관 수는 1,544기관으로 16년도 1,487기관에 비하여 57기관이 증가되었으며 평균 회신율을 88.3%였다.

2. 재료

요화학검사를 위한 관리물질은 1차(CU-17-01-04), 2차(CU-17-05-08), 3차(CU-17-09-12) 검체 모두 국내에서 제조한 액상관리물질 4종씩을 각 회원기관에 냉장 포장하여 발송하였다. 이 중 1차 및 3차에는 본 협회 홈페이지에 게시한 요침사 사진 8매 등 모두 20종의 검체를 우송 및 게시하여 정도관리 신빙도조사를 시행하였다. 결과치의 입력은 본 협회에서 구축한 신빙도조사사업 홈페이지에서 직접 입력하는 방식을 사용하였다. FOB를 위한 관리물질도 1차 및 2차 검체 모두 국내에서 제조한 점액성 분말관리물질 3종(CS-17-01-06)을 각 회원기관에 냉장 포장하여 발송하였다.

3. 정도관리물질의 허용치

요화학 및 FOB검사의 경우 분과위원회에서 실험한 참고치, 각 기관에서 회신한 결과의 일치도를 고려하였으며, 정성값은 최빈도치±1단계, 정량값은 평균±2 standard deviation의 범위를 허용범위로 설정하였다.

1. 요화학검사 9종의 신빙도조사 성적

요화학검사 9종(pH, protein, glucose, ketone, bilirubin, blood, urobilinogen, nitrite, leukocyte)의 신빙도조사 성적은 Table 1에 요약하였다. Table 1은 1차, 2차 및 3차 신빙도조사에서 얻어진 평균 정답률을 구한 자료이며, 모든 항목에서 95% 이상의 정답률을 나타내었다.

Table 1 . Number of participants and acceptable rates of urine chemistry QC.

   Tests1st QC2nd QC3rd QC



ParticipantsAcceptable (%)ParticipantsAcceptable (%)ParticipantsAcceptable (%)
pH1,31295.11,31097.31,45098.2
Protein1,31596.51,31597.51,45697.1
Glucose1,31298.81,31097.41,45198.1
Ketone1,28299.81,28199.61,41599.6
Bilirubin1,28299.11,28199.31,41599.2
Blood1,31297.91,31096.91,44998.6
Urobilinogen1,28299.61,28198.31,41599.6
Nitrite1,26699.31,26699.31,39399.7
Leukocyte1,25797.21,25798.71,38597.7

Abbreviation: QC, quality control..


2. 요화학 분석기에 따른 요화학검사 신빙도조사 성적

각 회사별 요화학 분석기에 따른 요화학검사의 신빙도조사 성적은 Table 2와 같다. YD (YD Diagnostics, Yongin, Korea)가 42.8%로 가장 많이 사용하고 있으며 SD (Standard Diagnostics, Seoul, Korea) 20.2%, Roche (Roche, Mannheim, Germany) 16.0%, DFI (DFI Co. Ltd., Gimhae, Korea) 9.3% 순이었다. 회사별 장비에 따라 요화학검사의 정답률은 차이를 보이지 않았으나 Beckman 장비 (Beckman Coulter Inc., Brea, CA, USA)는 백혈구 검출에서 75%의 낮은 정답률을 나타내었으며, 이는 백혈구 양성 검체를 음성으로 판독한 데서 기인하였다.

Table 2 . Acceptable rates according to urine analyzers from different companies.

 CompanypHProteinGlucoseKetoneBilirubinBloodUrobilinogenNitriteLeukocyteTotal
YD97.698.498.697.999.898.699.799.697.4545 (42.8)
SD96.897.198.999.199.896.899.399.098.6256 (20.2)
Roche94.198.799.898.6100.098.798.099.399.1203 (16.0)
DFI96.285.595.498.699.097.699.699.397.8118 (9.3)
Siemens99.999.799.299.896.698.798.499.899.280 (6.3)
Eiken95.799.599.399.091.899.598.799.899.252 (4.1)
Arkray100.098.6100.0100.0100.098.099.599.7100.031 (2.5)
Symex100.098.5100.098.097.9100.098.9100.092.614 (1.1)
Beckman100.097.799.2100.099.278.9100.0100.075.09 (0.7)

Values are presented as % or number (%)..

The instruments used were from the following companies: YD (YD Diagnostics, Yongin, Korea), SD (Standard Diagnostics, Seoul, Korea), Roche (Roche diagnostics, Mannheim, Germany), DFI (DFI Co. Ltd., Gimhae, Korea), Siemens (Siemens Healthcare (Erlangen, Germany), Eiken (Eiken Chemical Co., Tokyo, Japan), Arkray (Arkray Inc., Kyoto, Japan), and Beckman (Beckman Coulter Inc., Brea, CA, USA)..


Beckman 장비의 음성 판독은 2016년부터 지속되어 그 원인을 분석하고자 하였다[2]. 대구가톨릭대학병원 진단검사의학과에서는 환자들의 소변을 YD사의 Uriscan Super 장비로 분석하여 백혈구 양성으로 나온 132예를 대상으로, Beckman사의 iChem Velocity 장비로 재검하였다. 124예에서 1단계 이하의 일치를 보여 93.9%의 일치율을 나타내어, 환자 검체 분석에는 위음성 현상이 관찰되지 않았다. 이에 기질효과(matrix effect)가 의심되어 정도관리물질을 한국고분자 시험연구소(Seoul, Korea)에 의뢰하여 분석하였다. 분석결과 2-methyl-p-anisidine (4-methoxy-o-toluidine) 및 n-methyl-p-anisidine이 백혈구 시약의 diazonium salt과 반응하는 것으로 보고하였다. 즉 백혈구의 esterase 효소 대신 상기 두 가지 물질이 반응하여 양성 결과를 나타낸다고 하였다. 이에 정도관리물질을 Beckman사와 YD사의 요시험지봉에 반응시킨 후 육안관찰을 시도하였다(Fig. 1A). 백혈구 시약에서 Beckman사는 연한 살색으로 YD사는 연자주빛으로 발색되었다. 백혈구에 의한 Beckman사의 요시험지봉 발색반응은 Fig. 1B와 같이 자주빛으로 발색된다.

Figure 1.

(A) Gross examination of CU-17-04. Beckman (left) and YD (right) strips show results of urine chemistry reactions. At the leukocyte pads, the Beckman strip displays a slightly yellowish color, while the YD strip displays a pale purple color. (B) Color chart for leukocytes for the Beckman strip. Abbreviations: UBG, urobilinogen; Vit C, vitamin C; SG, specific gravity; WBC, white blood cell.


3. 요화학 분석기로 시행한 요비중검사 신빙도조사 성적

요비중검사는 굴절계 검사는 시행하지 않고 요화학 분석 장기로 측정하였다. Table 3에서와 같이 모든 검체에서 95% 이상의 정답률을 나타내었다. 그러나 결과입력 시 잘못된 값이 입력되는 경우가 발생하여 각 기관에서 결과입력 시 주의가 필요하였다.

Table 3 . Acceptable rates of the urine specific gravity test.

 SpecimenParticipantsMean±standard deviation95% Confidence intervalAcceptable (%)
CU-17-011,2611.005±0.0281.001 to 1.01097.5
CU-17-021,2611.009±0.0041.005 to 1.01597.0
CU-17-031,2611.008±0.0281.005 to 1.01895.6
CU-17-041,2611.021±0.0061.010 to 1.03098.2
CU-17-051,2631.005±0.0281.001 to 1.01594.9
CU-17-061,2621.008±0.0041.005 to 1.01597.8
CU-17-071,2621.008±0.0041.005 to 1.01596.3
CU-17-081,2621.019±0.0401.010 to 1.03097.9
CU-17-091,3881.006±0.0271.001 to 1.01596.8
CU-17-101,3881.007±0.0271.005 to 1.01596.7
CU-17-111,3881.006±0.0471.005 to 1.01595.6
CU-17-121,3881.018±0.0541.010 to 1.03097.5

4. 사진 검체를 이용한 요침사물 검사결과

사진 검체를 사용하여 시행한 요침사물 신빙도조사의 평균 정답률은 68.67%에서 99.3%의 정답률을 보였다(Figs. 2, 3, Table 4). 지방구(CUI-17-02)를 전분입자나 백혈구로 판정하였고, 비정형 요산결정(CUI-17-05)을 과립원주나 적혈구 원주로 판정하였다. 또한 납양원주(CUI-17-07)를 초자원주로 판정하였고, 이행상피세포와 세뇨관상피세포(CUI-17-06)의 구별이 용이하지 않았다. 따라서 요침사물에 대한 판독은 각 기관의 지속적인 개선 노력이 필요하였다.

Figure 2.

Images of urine sediments (with arrow) for the external quality assessment. (A) CUI-17-01 uric acid crystal (original magnification ×400). (B) CUI-17-02 fat globule (original magnification ×100). (C) Polarized microscopic images of panel B. (D) CUI-17-03 cystine crystal (original magnification ×400). (E) CUI-17-04 cellular cast (original magnification ×400).


Figure 3.

Images of urine sediments (with arrow) for the external quality assessment. (A) CUI-17-05 uric acid crystal (original magnification ×200). (B) CUI-17-06 renal tubular epithelial cell (original magnification ×400). (C) CUI-17-07 waxy cast (original magnification ×100). (D) CUI-17-08 ammonium urate crystal (original magnification ×400).


Table 4 . Acceptable rates for the urine sediment examination.

 Specimen no.   ResultsParticipantsAnswersAcceptable (%)Unacceptable (%)
CUI-17-01Uric acid75274799.3
CUI-17-02Fat droplets75251668.6
CUI-17-02Starch granule752739.7
CUI-17-02Leukocyte752729.6
CUI-17-03Cystine75274498.9
CUI-17-04Granular cast coarsely75252369.5
CUI-17-04RBC/pigmented/muddy brown casts75212316.4
CUI-17-04Granular cast finely752618.1
CUI-17-04Cellular cast (RTE and/or neutrophil)752172.3
CUI-17-04Waxy cast752141.9
CUI-17-04Fatty cast752121.6
CUI-17-05Uric acid83866078.8
CUI-17-05Granular cast coarsely838404.8
CUI-17-05RBC/pigmented/muddy brown casts838404.8
CUI-17-05Fiber/fecal contamination838253.0
CUI-17-05Calcium oxalate838192.3
CUI-17-05Fatty cast838131.6
CUI-17-06Transitional epithelial cell83846855.8
CUI-17-06Renal tubular epithelial cell83833640.1
CUI-17-06Squamous epithelial cell838172.0
CUI-17-07Waxy cast83860372.0
CUI-17-07Fiber/fecal contamination83810512.5
CUI-17-07Hyaline cast838738.7
CUI-17-07Granular cast finely838273.2
CUI-17-07Cholesterol838192.3
CUI-17-08Ammonium biurate83879695.0
CUI-17-08Amorphous urates/phosphates838232.7
CUI-17-08Leucine838141.7

5. FOB 시약에 따른 FOB 정성검사의 정답률 및 시약의 분포

FOB 시약에 따른 FOB 정성검사의 정답률은 Table 5에 나타내었다. 시약은 SD FOB rapid test (268기관), YD FOB test (121기관), Eiken homocatch light (Eiken Chemical Co., Tokyo, Japan; 56기관), Asan FOB easy test (Asan Pharmaceutical, Seoul, Korea; 52기관) 순으로 사용하였다. 모든 시약에서 양호한 정답률을 나타내었으나, FOB 음성 검체에서 SD FOB, Humasis FOB test (Humasis Co., Anyang, Korea) 및 GC Genedia FOB (Green Cross Medial Science Co., Yongin, Korea) 시약에서 90% 이하의 낮은 정답률을 나타내었다. 특히 Humasis사 시약은 음성 검체 2개에서 모두 90% 이하의 정답률을 나타내어 시약의 개선이 필요하였다.

Table 5 . Acceptable rates for the fecal occult blood quality test.

   ReagentsNo.CS-16-01CS-16-02CS-16-03CS-16-04CS-16-05CS-16-06

Negative (%)Positive (%)Positive (%)Positive (%)Negative (%)Positive (%)
SD FOB rapid test26892.999.6100.0100.086.2100.0
YD Occult Tech FOB test12194.4100.0100.0100.090.199.2
Eiken Hemocatch light56100.0100.098.2100.0100.0100.0
ASAN easy test FOB5292.7100.0100.0100.096.298.1
Humasis FOB test3984.6100.0100.0100.082.1100.0
Bio focus FOB rapid test3093.3100.0100.0100.093.3100.0
HBI HiSens FOB NCard2696.796.796.7100.096.2100.0
GC Genedia FOB1894.194.1100.0100.088.9100.0

Abbreviation: FOB, fecal occult blood..

The instruments used were from the following companies: SD (Standard Diagnostics, Seoul, Korea), YD (YD Diagnostics, Yongin, Korea), Eiken Chemical Co. (Tokyo, Japan), Asan Pharmaceutical (Seoul, Korea), Humasis Co. (Anyang, Korea), Bio Focus Co. (Uiwang, Korea), HBI Co. (Anyang, Korea), and GC (Green Cross Medial Science Co., Yongin, Korea)..


6. FOB 시약에 따른 FOB 정량검사의 정답률 및 시약의 분포

FOB 정량검사결과 분포는 Table 6에 요약하였다. 장비별로 검사에 사용하는 대변 양이 달라서 장비 간 비교가 어려워 이를 보정한 값을 함께 비교하였다. 사용장비는 Eiken (158기관), Alfresa (Alfresa Pharma Co., Osaka, Japan; 60기관), Kyowa (Kyowa Chemical Industry Co., Kagawa, Japan; 35기관) 순으로 분포하였다. 전체 평균에 비하여 Alfresa 장비는 CS-17-01 및 CS-17-05 검체에서 높은 수치를 보였다. 또한 Alfresa 장비는 CS-17-02, 17-03, 17-04 및 17-06 검체에서 모두 낮은 농도를 나타내었다.

Table 6 . Distribution of results of the fecal occult blood quantitation test.

InstrumentsSpecimenNo.MedianMinimumMaximumAverage




ng/mLμg/g stoolng/mLμg/g stoolng/mLμg/g stoolng/mLμg/g stool
 EikenCS-17-011574.00.80.00.024.04.84.81.0
 AlfresaCS-17-015856.010.63.00.6149.028.353.710.2
 KyowaCS-17-01331.01.00.00.055.055.02.52.5
 EikenCS-17-02157195.039.082.016.4342.068.4195.439.1
 AlfresaCS-17-0258151.028.753.010.1213.040.5150.428.6
 KyowaCS-17-023348.048.00.00.096.096.048.648.6
 EikenCS-17-03157390.078.065.013.0561.0112.2379.976.0
 AlfresaCS-17-0358264.550.358.011.0370.070.3259.749.3
 KyowaCS-17-0333107.0107.068.068.0165.0165.0103.9103.9
 EikenCS-17-04158465.593.154.010.8620.0124.0450.490.1
 AlfresaCS-17-0460302.057.451.09.7400.076.0295.756.2
 KyowaCS-17-0435133.0133.075.075.0181.0181.0127.1127.1
 EikenCS-17-051584.00.80.00.018.03.64.00.8
 AlfresaCS-17-056046.08.71.00.2122.023.254.510.4
 KyowaCS-17-05351.01.00.00.07.07.01.11.1
 EikenCS-17-06158232.046.4107.021.4420.084.0226.045.2
 AlfresaCS-17-0660179.034.0111.021.1325.061.8180.534.3
 KyowaCS-17-063557.057.022.022.073.073.053.653.6

The instruments used were from the following companies: Eiken Chemical Co. (Tokyo, Japan), Alfresa Pharma Co. (Osaka, Japan), and Kyowa Chemical Industry Co. (Kagawa, Japan)..


2017년의 정도관리 신빙도조사에서 나타난 특징은 참여기관의 증가이다. 특히 3차 신빙도조사에서는 1,450기관이 참가하여 1차에 비하여 10.7%의 증가를 보였다. 결과를 분석해 볼 때 2016년 정도관리 신빙도사업결과와 매우 유사한 성적을 나타내었다[2]. 백혈구를 제외한 요화학검사 및 비중검사는 우수한 정답률을 유지하였다. 요침사 신빙도조사에서도 2016년 정답률 67.8%-86.7%보다 조금 상향된 성적을 보였다. FOB 정성검사에서는 2016년과 유사하게 음성 검체에 대한 위양성률이 높지 않아 국내 시약의 질 개선이 이루어진 것으로 보인다.

요화학검사에서는 2016년과 동일하게 Beckman 장비에서 leukocytes 양성인 검체를 음성으로 판독하였다. 이에 소변, 대변검사 위원회와 Beckman사가 공동으로 원인분석을 시행하였다. Beckman사 장비는 실제 환자들의 소변 검체에서는 위음성을 보이지 않았으나, 정도관리 검체에서는 위음성을 보여 기질효과를 의심하였다. 한국고분자 시험연구소에서 정도관리물질 분석을 의뢰한 결과 백혈구 시약반응에 양성을 일으킬 수 있는 2가지 물질을 발견하였다. Beckman사 시약과 YD사 시약을 대상으로 정도관리물질을 반응시켜 육안 판독한 결과, Beckman사 시약에서는 발색반응이 비특이적으로 일어났다. 즉 백혈구의 esterase과 반응하여 발색되는 자색반응이 아니라 살색반응이 나타나서 장비에서 잘 검출하지 못하는 것으로 판단되었다. 따라서 Beckman사 장비로 신빙도조사를 시행할 경우 장비판독과 육안판독을 모두 시행하여, 비특이적인 발색반응으로 인한 위음성 반응이 의심되는 경우, 육안판독 결과를 보고하는 것이 바람직할 것이다.

요비중검사는 참여기관의 정답률이 양호한데, 드물게 결과 입력오류가 발견되었다. 0, 0.01, 0.02 등 비정상적인 결과가 입력되는데, 입력방법을 잘 숙지하여 입력오류에 의한 불합격판정을 받지 않아야겠다.

요침사 신빙도조사에서는 지방구, 비정형 요산결정 등에서 낮은 정답률을 기록하였다. 두 가지 침사 모두 빈번하게 나타나는 형태는 아니지만 각각 신증후군과[3] 요산결석을[4] 시사할 수 있다. 정답으로 평가하였지만 세뇨관 상피세포와 이행상피세포의 감별은 아직도 잘 이루어지지 못하고 있다. 요침사물을 정확하게 판독하려면 판독경험이 많아야 하므로 각 병원에서 요침사 판독전문가를 1명 이상 육성하는 것이 바람직할 것이다.

FOB 정성검사에서 2015년 전 시약에서 음성 검체를 양성으로 판독하여 오답률이 89.0%에 달하여 문제점으로 부각되었다. 정성시약의 과도한 민감도를 낮추기 위하여 국내 FOB 시약회사들에게 민감도 조정을 요청하였고, 2016년에 2회에 걸쳐 FOB 음성인 검체로 신빙도조사사업을 시행한 결과 모두 90% 이상의 정답률을 보였다. 2017년에도 FOB 음성인 검체에서 84.6% 이상의 정답률을 보여 다소 정답률이 하락하였으나, 위양성 문제는 해결된 것으로 보였다. 본 위원회에서는 시판되는 6개 시약(SD, YD, Eiken, Asan, Humasis 및 BioFocus)을 대상으로 2017년 11월 민감도 개선 여부를 확인하였으며, 20 ng/mL 이하의 환자 대변 검체에서 평균 양성률 6.4%를 나타내었다(data not shown). 이 결과에 대한 세부내용은 논문으로 출판할 것이다.

대변잠혈 정량검사에서 Alfresa는 2016년과 동일하게 저농도 검체에서는 높게, 중등 및 고농도 검체에서는 낮게 측정하였다. Kyowa 장비는 중등 및 고농도 검체에서 다소 높게 측정되었다. FOB는 장비에 따라 결과 차이를 보일 수 있으며[5], 특히 측정 시 주입되는 대변 양이 달라서 ng/mL의 단위로 보고하면 농도차이를 보일 수 있다. 이를 보완하기 위하여 FOB의 또 다른 보고단위인 μg/g stool을 도입하여 병행 사용을 추진하였으나[6], 참가기관에서 주입하는 대변 양을 정확하게 측정하기 어려워 2017년 신빙도조사사업에서 시행하지 못하였다. Table 6에 제시된 μg/g stool에 의한 수치는 실제치가 아니고 장비별로 주입되는 대변 양을 일괄적으로 적용하여 산출된 수치이므로 정확하지 않다. 따라서 장비별로 발생하는 FOB 결과차이에 대한 원인을 분석하기는 어려우며, 사용하는 기관에서 장비별 특성을 잘 고려하여 참고치를 산정해야 할 것이다.

  1. Jeon CH, Lee AJ. Urinalysis and Routine Microscopy Subcommittee, Korean Association of External Quality Assessment Service. Annual report on the external quality assessment scheme for urinalysis and faecal occult blood testing in Korea (2015). J Lab Med Qual Assur 2016;38:120-8.
    CrossRef
  2. Jeon CH, Lee AJ, Kim SG, Suh HS, Bae YC. Annual report on the external quality assessment scheme for urinalysis and faecal occult blood testing in Korea (2016). J Lab Med Qual Assur 2017;39:117-23.
    CrossRef
  3. McPherson RA, Pincus MR. Henry's clinical diagnosis and management by laboratory methods. 23rd ed. St. Louis (MO): Elsevier, 2017.
  4. Cicerello E. Uric acid nephrolithiasis: an update. Urologia 2018;85:93-8.
    Pubmed CrossRef
  5. Kim JH, Hwang SY, Kim YJ. Evaluation of Hemo Techt NS-plus C15 Automatic Analyzer for fecal occult blood test. J Lab Med Qual Assur 2010;32:237-41.
  6. Sinatra MA, Young GP, St John DJ, Blake D, Ratnaike S. A study of laboratory based faecal occult blood testing in Melbourne, Australia. The Faecal Occult Blood Testing Study Group. J Gastroenterol Hepatol 1998;13:396-400.
    Pubmed CrossRef

Supplementary File