Laboratory Medicine and Quality Assurance

Table. 1.

Accuracy of the two infusion systems

Sample no.Infusion rate (mL/hr)Time (min)TerumoBaxter


Infused volume (mL)Error rate (%)Infused volume (mL)Error rate (%)
11018030.10.333.712.3
21018032.37.733.612.0
31018031.13.734.113.7
43012062.54.272.520.8
53012063.15.268.213.7
63012062.54.265.08.3
710060104.64.6110.010.0
810060102.02.0107.97.9
910060101.01.0112.912.9

The instruments were from the following companies: Baxter Health-care Corp. (Deerfield, IL, USA) and Terumo Medical Corp. (Tokyo, Japan).

J Lab Med Qual Assur 2019;41:161~165 https://doi.org/10.15263/jlmqa.2019.41.3.161