Laboratory Medicine and Quality Assurance

Table. 7.

Results of proficiency tests for quantitative anti-TPO in 2018 and 2019

Trial Manufacturer No. of partici­pants (%) No. of out­liers Mean±SD CV (%) Median Min Max
2018, 1st Total 61 (100.0) 9 146.78±24.31 16.6 144.5 77.2 370.3
Abbott Diagnostics 10 (16.4) 181.97±11.78 6.5 180.4 168.7 199.0
Beckman Coulter 4 (6.6) 120.75±1.62 1.3 121.1 118.7 122.1
Diesse Diagnostica Senese SpA 2 (3.3) 241.15±73.75 30.6 189.0 293.3
Grifols 1 (1.6) 146.00 146.0
Roche Diagnostics 34 (55.7) 138.99±8.30 6.0 139.9 124.0 199.2
Siemens Healthineers 9 (14.8) 355.33±13.76 3.9 358.9 77.2 370.3
Others 1 (1.6) 132.00 132.0
2018, 2nd Total 64 (100.0) 9 75.95±14.10 18.6 80.4 41.0 394.0
Abbott Diagnostics 11 (17.2) 66.47±3.59 5.4 67.0 60.3 72.6
Beckman Coulter 4 (6.3) 44.10±2.72 6.2 44.0 41.0 47.4
Diesse Diagnostica Senese SpA 2 (3.1) 77.95±38.25 49.1 50.9 105.0
Dynex Technologies 1 (1.6) 92.00 92.0
Grifols 1 (1.6) 89.60 89.6
Roche Diagnostics 37 (57.8) 81.45±8.50 10.4 82.3 61.5 394.0
Siemens Healthineers 8 (12.5) 207.89±16.42 7.9 203.7 187.4 232.1
2019, 1st Total 80 (100.0) 11 18.16±3.81 21 19.7 9.7 114.2
Abbott Diagnostics 13 (16.3) 18.10±1.33 7.4 17.6 15.6 19.7
Beckman Coulter 5 (6.3) 12.33±0.12 0.9 12.4 11.7 13.3
bioMérieux 1 (1.3) 9.70 9.7
DiaSorin 1 (1.3) 28.60 28.6
Diesse Diagnostica Senese SpA 1(1.3) 17.60 17.6
Dynex Technologies 1 (1.3) 42.60 42.6
Grifols 1 (1.3) 31.20 31.2
Roche Diagnostics 48 (60.0) 18.78±3.61 19.2 19.9 11.8 114.2
Siemens Healthineers 9 (11.3) 77.66±5.54 7.1 76.8 17.1 91.7
2019, 2nd Total 81 (100.0) 9 120.58±22.33 18.5 115.0 2.5 897.3
Abbott Diagnostics 14 (17.3) 157.41±8.31 5.3 159.6 144.2 170.6
Beckman Coulter 6 (7.4) 106.06±4.07 3.8 105.2 2.5 110.1
bioMérieux 1 (1.2) 97.60 97.6
DiaSorin 1 (1.2) 167.00 167.0
Dynex Technologies 1 (1.2) 132.00 132.0
Grifols 1 (1.2) 154.00 154.0
Roche Diagnostics 48 (59.3) 109.44±9.19 8.4 111.9 91.5 131.3
Siemens Healthineers 9 (11.1) 780.85±78.07 10.0 737.9 144.0 897.3

Manufacturers of instrument are listed in alphabetical order. The instruments were from the following companies: Abbott Diagnostics (Abbott Park, IL, USA); Beckman Coulter (Brea, CA, USA); Diesse Diagnostica Senese SpA (Monteriggioni, Italy); Grifols (Barcelona, Spain); Roche Diagnostics (Mannheim, Germany); Siemens Healthineers (Erlangen, Germany); Dynex Technologies (Chantilly, VA, USA); bioMérieux (bioMérieux, Marcy-l'Etoile, France); and DiaSorin (DiaSorin, Saluggia, Italy).

Abbreviations: anti-TPO, anti-thyroperoxidase antibody; SD, standard deviation; CV, coefficient of variation.

Lab Med Qual Assur 2020;42:1~9 https://doi.org/10.15263/jlmqa.2020.42.1.1