Laboratory Medicine and Quality Assurance

Table. 3.

Distribution of instruments and mean, SD, and CVs (%) for glucose point-of-care testing (GGP-18-01 to GGP-18-03)

Manufacturer Instrument No. GGP-18-01 GGP-18-02 GGP-18-03



Mean±SD CV (%) Mean±SD CV (%) Mean±SD CV (%)
Abbott Freestyle Optimum 32 78.3±2.5 3.2 143±6.9 4.8 280.3±11.5 4.1
Freestyle Optimum Neo 22 66.2±2.1 3.1 152±6.1 4.0 356.3±22.5 6.3
Freestyle Optimum Neo H 7
Freestyle Optimum Xceed 49 78.5±4.2 5.3 149.4±5.6 3.7 293.2±17.3 5.9
Allmedicus GlucoDr. Auto (AGM-4000) 4
GlucoDr. Plus (AGM-3000/B) 101 44.5±2.5 5.7 105.7±4.4 3.9 254.7±9.4 3.7
GlucoDr. (AGM-2100) 1
Others
Arkray Inc. Glucocard 01 14 41.3±6.0 14.5 97.9±12.9 13.2 290.1±31.5 10.9
Glucocard Ⅱ 2
Ascensia Breeze 2 18 39.7±0.8 1.9 116.4±4.6 4.0 331.7±8.8 2.7
Contour plus 38 55.6±2.7 4.8 111.2±3.8 3.4 263.7±5.1 1.9
Contour T 19 31.7±1.6 4.9 94.9±6.6 7.0 352.7±11.0 3.1
GC Medis Green check
Others
Infopia GlucoLAB 13 53.0±0.0 0.0 112.3±2.3 2.1 286.3±5.2 1.8
GlucoLAB Auto-coding 7
Gluneo Lite 1
Gluneo Plus 118 55.7±4.9 8.8 88.1±5.3 6.0 226.5±7.7 3.4
Mirae 3.3G+ 5
Others 3
i-SENS ACURA plus
BAROZen 52 38.1±3.1 8.2 96.6±6.2 6.4 263.6±35.7 13.6
BAROZen H 147 41.0±7.1 17.3 104.3±6.1 5.8 304.1±23.1 7.6
BAROzen H expert 317 52.4±3.7 7.0 116.3±7.0 6.0 330.5±14.5 4.4
BAROzen H expert plus 81 53.0±1.8 3.4 112.3±3.5 3.1 327.9±9.1 2.8
BAROZen H plus 129 53.3±2.5 4.8 118.0±5.8 4.9 339.1±16.1 4.7
BAROZen II
BAROZen plus
CareSens Dual
CareSens II 15 32.4±2.1 6.6 96.1±6.3 6.6 259.0±21.7 8.4
CareSens N 80 39.0±3.2 8.2 99.0±8.0 8.1 276.6±45.9 16.6
CareSens N Voice 103 41.0±4.2 10.2 103.3±5.5 5.3 291.9±17.7 6.1
CareSens PRO 78 46.5±3.4 7.2 110.1±5.7 5.2 376.7±48.3 12.8
Others 69 38.2±2.6 6.9 99.6±6.8 6.8 282.4±23.0 8.1
Philosys Gmate check
Gmate Origin
Gmate WHeeL
Roche Accu-Check Active 198 78.3±2.3 2.9 165.5±4.5 2.7 419.2±11.1 2.6
Accu-Check Inform II 411 68.5±1.5 2.2 157.4±3.4 2.2 406.9±9.5 2.3
Accu-Check Performa 619 68.8±1.3 2.0 158.4±3.4 2.2 408.7±7.4 1.8
Others
SD Biosensor SD CHECK GOLD (Meter 01GM10) 36 63.7±1.7 2.6 139.0±3.6 2.6 375.4±9.1 2.4
SD CHECK GOLD (System 01GC12) 7
SD CodeFree (Meter 01GM11) 117 55.6±5.3 9.5 133.7±7.1 5.3 353.9±11.3 3.2
SD CodeFree (System 01GC112) 8
SD GlucoNavii
SD GlucoNavii GDH (Meter 01GM30)
SD GlucoNavii GDH (System 01GC32)
SD GlucoNavii Link 0.3 (Meter 01GM12) 40 58.5±2.7 4.7 145.0±4.9 3.4 396.2±17.9 4.5
SD GlucoNavii Link 0.3 (System 01GC122) 12 54.7±2.0 3.6 130.8±3.2 2.4 354.0±4.3 1.2
SD GlucoNavii NFC (Meter 01GM40) 78 62.1±4.7 7.6 136.1±11.7 8.6 325.1±32.5 10.0
SD GlucoNavii NFC (System 01GC42) 6
Wise Meditech Wisecheck
Others Others 98 47.0±15.5 33.1 106.5±17.1 16.1 257.7±6.9 2.7

The instruments used were from the following companies: Abbott (Abbott Park, IL, USA), Allmedicus (Anyang, Korea), Arkray Inc. (St. Edina, MN, USA), Ascensia (Newbury, UK), GC Medis (Cheonan, Korea), Infopia (Anyang, Korea), i-SENS (Seoul, Korea), Philosys (Gunsan, Korea), Roche (Indianapolis, IN, USA), SD Biosensor (Suwon, Korea), and Wise Meditech (Anyang, Korea).

Abbreviation: SD, standard deviation; CV, coefficient of variance.

Lab Med Qual Assur 2021;43:43~55 https://doi.org/10.15263/jlmqa.2021.43.2.43
© Lab Med Qual Assur