Table. 4.

Distribution of instruments and mean, SD, and CVs (%) for glucose point-of-care testing (GGP-18-04 to GGP-18-06)

Manufacturer Instrument group No. GGP-18-04 GGP-18-05 GGP-18-06



Mean±SD CV (%) Mean±SD CV (%) Mean±SD CV (%)
Abbott Freestyle Optimum 33 74.6±2.3 3.1 151.4±6.5 4.3 317.2±19.7 6.2
Freestyle Optimum Neo 22 67.3±4.4 6.6 164.8±12.4 7.5 354.7±27.9 7.9
Freestyle Optimum Neo H 7
Freestyle Optimum Xceed 24 79.0±5.3 6.7 161.1±6.0 3.8 321.0±18.7 5.8
Allmedicus GlucoDr. Auto (AGM-4000) 4
GlucoDr. Plus (AGM-3000/B) 68 46.3±3.2 6.9 114.0±4.4 3.9 264.5±11.5 4.3
GlucoDr. (AGM-2100) 1
Others
Arkray Inc Glucocard 01 9
Glucocard Ⅱ 2
Ascensia Breeze 2
Contour plus 43 55.4±2.1 3.8 116.3±4.4 3.7 270.3±7.3 2.7
Contour T 23 41.3±13.7 33.2 107.7±11.4 10.6 321.1±44.1 13.7
GC Medis Green check
Others
Infopia GlucoLAB 17 53.0±1.7 3.2 117.9±2.0 1.7 258.9±8.1 2.8
GlucoLAB Auto-coding 20 52.4±7.0 13.4 119.5±8.2 6.9 296.9±14.4 4.9
Gluneo Lite
Gluneo Plus 133 57.5±3.4 5.8 96.0±3.2 3.4 236.9±6.5 2.7
Mirae 3.3G+ 5
Others
i-SENS ACURA plus
BAROZen 66 40.1±2.9 7.2 106.9±6.2 5.8 284.2±25.7 9.1
BAROZen H 114 39.9±2.1 5.2 111.9±4.0 3.6 315.0±12.2 3.9
BAROzen H expert 421 55.8±3.8 6.8 120.6±5.7 4.8 326.6±14.9 4.6
BAROzen H expert plus 83 54.5±3.6 6.6 119.5±5.0 4.2 326.6±13.8 4.2
BAROZen H plus 145 54.2±3.6 6.6 119.6±5.0 4.1 327.2±16.5 5.1
BAROZen II
BAROZen plus
CareSens Dual
CareSens II 11 32.7±3.1 9.6 98.2±3.8 3.9 282.7±11.1 3.9
CareSens N 83 41.4±3.9 9.5 109.8±5.2 4.7 302.1±20.0 6.6
CareSens N Voice 118 38.9±3.3 8.5 106.7±4.5 4.2 302.1±16.3 5.4
CareSens PRO 81 51.7±3.8 7.3 119.8±5.2 4.3 333.7±18.8 5.6
Others 50 36.7±4.2 11.3 107.2±4.4 4.1 290.8±23.7 8.2
Philosys Gmate check
Gmate Origin
Gmate WHeeL
Roche Accu-Check Active 178 77.9±2.7 3.5 172.2±5.8 3.4 428.1±9.4 2.2
Accu-Check Inform II 413 69.5±1.4 2.0 166.5±3.1 1.9 403.6±8.3 2.1
Accu-Check Performa 673 70.0±1.5 2.1 166.2±3.4 2.0 403.1±8.3 2.1
Others
SD Biosensor SD CHECK GOLD (Meter 01GM10) 37 60.2±3.8 6.3 143.2±8.9 6.2 372.1±17.3 4.6
SD CHECK GOLD (System 01GC12) 6
SD CodeFree (Meter 01GM11) 87 58.1±5.7 9.8 146.3±6.6 4.5 368.1±11.8 3.2
SD CodeFree (System 01GC112) 10 54.8±1.6 2.9 137.3±2.5 1.9 353.3±2.2 0.6
SD GlucoNavii
SD GlucoNavii GDH (Meter 01GM30)
SD GlucoNavii GDH (System 01GC32)
SD GlucoNavii Link 0.3 (Meter 01GM12) 40 58.6±4.6 7.8 142.4±7.2 5.1 373.2±13.1 3.5
SD GlucoNavii Link 0.3 (System 01GC122) 12 57.8±4.8 8.3 140.6±8.2 5.8 337.6±14.4 4.3
SD GlucoNavii NFC (Meter 01GM40) 82 64.1±3.6 5.6 144.9±9.2 6.3 348.8±33.7 9.7
SD GlucoNavii NFC (System 01GC42) 11 61.5±1.3 2.1 142.8±2.1 1.5 349.3±7.7 2.2
Wise Meditech Wisecheck
Others Others 130 43.5±9.1 20.8 113.2±12.3 10.9 269.3±36.1 12.2

The instruments used were from the following companies: Abbott (Abbott Park, IL, USA), Allmedicus (Anyang, Korea), Arkray Inc. (St. Edina, MN, USA), Ascensia (Newbury, UK), GC Medis (Cheonan, Korea), Infopia (Anyang, Korea), i-SENS (Seoul, Korea), Philosys (Gunsan, Korea), Roche (Indianapolis, IN, USA), SD Biosensor (Suwon, Korea), and Wise Meditech (Anyang, Korea).

Abbreviation: SD, standard deviation; CV, coefficient of variance.

Lab Med Qual Assur 2021;43:43~55 https://doi.org/10.15263/jlmqa.2021.43.2.43
© Lab Med Qual Assur