Laboratory Medicine and Quality Assurance

Table. 2.

Between-assay correlations of measurement results for patient samples

Analyte Assay pair (x–y) Slope Intercept
Alpha-fetoprotein Roche-Abbott 0.8988 (0.8808 to 0.9135) 1.7460 (0.3642 to 3.4328)
Roche-Siemens 0.8924 (0.8776 to 0.9084) 0.2542 (–0.9436 to 0.9050)
Roche-Beckman 0.9945 (0.9780 to 1.0098) –0.0410 (–1.4764 to 0.5896)
Abbott-Siemens 0.9983 (0.9769 to 1.0179) –1.1876 (–3.2560 to 0.0755)
Abbott-Beckman 1.1246 (1.1004 to 1.1506) –2.1669 (–5.7047 to –0.2924)
Siemens-Beckman 1.1119 (1.0941 to 1.1374) –0.6813 (–1.8866 to 0.1566)
Carcinoembryonic antigen Roche-Abbott 1.0869 (1.0193 to 1.1582) –1.0863 (–2.9313 to –0.2211)
Roche-Siemens 0.7309 (0.6847 to 0.7937) –0.6278 (–2.3317 to 0.1883)
Roche-Beckman 1.1696 (1.0833 to 1.2676) –1.1795 (–3.7368 to 0.0173)
Abbott-Siemens 0.6548 (0.5783 to 0.7256) 0.3653 (–0.4564 to 1.6287)
Abbott-Beckman 1.1038 (1.0091 to 1.2507) –0.3280 (–2.2638 to 0.8486)
Siemens-Beckman 1.6587 (1.5705 to 1.7286) –0.7319 (–1.5417 to 0.2363)
Prostate-specific antigen Roche-Abbott 0.9324 (0.9115 to 0.9495) 0.0517 (–0.1001 to 0.2945)
Roche-Siemens 0.7906 (0.7659 to 0.8165) 0.0306 (–0.3014 to 0.2181)
Roche-Beckman 1.1765 (1.1459 to 1.2034) –0.0687 (–0.3715 to 0.2397)
Abbott-Siemens 0.8465 (0.8167 to 0.8685) –0.0983 (–0.3507 to 0.2197)
Abbott-Beckman 1.2535 (1.2206 to 1.2793) –0.2107 (–0.4366 to 0.1923)
Siemens-Beckman 1.4798 (1.4378 to 1.5259) –0.0133 (–0.3760 to 0.4964)

Values are presented as Pearson correlation coefficients (95% confidence interval). The instruments used were from Abbott Laboratories (Abbott Park, IL, USA), Beckman Coulter Inc. (Brea, CA, USA), Roche Diagnostics (Mannheim, Germany), and Siemens Healthineers (Erlangen, Germany)

Lab Med Qual Assur 2022;44:111~120 https://doi.org/10.15263/jlmqa.2022.44.2.111