Laboratory Medicine and Quality Assurance

Table. 3.

Frozen human serum pools concentrations and measured CVs with different assays

Analyte Assay Median (range) (ng/mL) Total CV (%)
Alpha-fetoprotein Abbott 87.10 (17.96–164.90) 1.8
Beckman 97.25 (20.51–186.44) 4.4
Roche 97.67 (18.60–185.33) 1.6
Siemens 87.00 (16.67–158.97) 1.2
Carcinoembryonic antigen Abbott 13.85 (3.44–35.35) 2.3
Beckman 13.36 (3.90–35.57) 5.9
Roche 13.63 (3.74–33.80) 2.0
Siemens 7.99 (2.90–20.79) 2.8
Prostate-specific antigen Abbott 45.46 (11.73–96.83) 2.4
Beckman 61.72 (13.95–127.84) 3.1
Roche 48.73 (13.10–108.00) 1.8
Siemens 39.39 (10.25–86.61) 2.0

The instruments used were from Abbott Laboratories (Abbott Park, IL, USA), Beckman Coulter Inc. (Brea, CA, USA), Roche Diagnostics (Mannheim, Germany), and Siemens Healthineers (Erlangen, Germany)

Abbreviation: CV, coefficient of variation

Lab Med Qual Assur 2022;44:111~120 https://doi.org/10.15263/jlmqa.2022.44.2.111