Laboratory Medicine and Quality Assurance

Table. 6.

Acceptable rates in the FOB quality test

Reagents Specimen (intended response) CS-21-01 (Pos) CS-21-02 (Neg) CS-21-03 (Pos) Specimen (intended response) CS-21-04 (Neg) CS-21-05 (Neg) CS-21-06 (Pos)
Abbott Bioline FOB 343 99.7 86.9 98.5 350 93.4 99.7 99.7
YD Occult Tech FOB 158 100.0 98.1 93.0 157 98.1 100.0 99.4
ASAN Easy Test FOB 66 100.0 83.3 97.0 70 95.7 98.6 98.6
Humasis FOB test 41 100.0 95.1 97.6 41 97.6 100.0 100.0
Eiken Hemocatch 31 100.0 100.0 100.0 32 100.0 100.0 100.0
GC Genedia FOB 25 100.0 84.0 92.0 25 96.0 100.0 100.0
Bio Focus FOB Rapid 10 100.0 70.0 90.0 10 90.0 100.0 100.0

Values are presented as the number of intended responses or acceptable percentages, unless otherwise stated. The results with <90% accuracy are indicated in bold. The instruments used were as follows: Abbott (Abbott Laboratories, Abbott Park, IL, USA), YD (YD Diagnostics, Yongin, Korea), Asan (Asan Pharmaceutical, Seoul, Korea), Humasis (Humasis Co., Anyang, Korea), Eiken (Eiken Chemical Co., Tokyo, Japan), GC (Green Cross Medial Science Co., Yongin, Korea), and Bio Focus (Bio Focus Co., Uiwang, Korea).

Abbreviations: FOB, fecal occult blood; Neg, negative; Pos, positive.

Lab Med Qual Assur 2023;45:1~9 https://doi.org/10.15263/jlmqa.2023.45.1.1