Laboratory Medicine and Quality Assurance

Table. 7.

Distribution of results of fecal occult blood quantitative test

Specimen no. Instruments No. of participants Mean±SD CV Minimum Maximum
CS-21-01 All 316 257.3±23.4 9.1 28 1,000
Eiken Oc sensor 184 259.9±15.3 5.9 53 510
Alfresa NS-Prime 54 255.2±21.2 8.3 37 418
Alfresa NS-Plus C 13 220.9±17.3 7.8 191 308
Kyowa HM-Jack 35 83.1±13.1 15.8 28 102
SD Standard 22 529.0±124.6 23.6 288 777
CS-21-02 All 316 5.7±7.9 138.5 0 83
Eiken Oc sensor 184 1.4±2.0 141.3 0 58
Alfresa NS-Prime 54 12.9±4.3 33.4 4 26
Alfresa NS-Plus C 13 61.5±15.3 24.8 34 83
Kyowa HM-Jack 35 0.5±0.6 112.5 0 2
SD Standard 22 25.0±0 0 0 31
CS-21-03 All 316 63.4±10.2 16.2 9 450
Eiken Oc sensor 184 61.6±8.1 13.1 24 264
Alfresa NS-Prime 54 67.6±8.0 11.9 12 99
Alfresa NS-Plus C 13 83.6±13.2 15.7 65 110
Kyowa HM-Jack 35 24.9±4.7 19.0 9 37
SD Standard 22 296.0±57.4 19.4 52 450
CS-21-04 All 316 1.3±2.4 184.6 0 66
Eiken Oc sensor 183 0 0 0 27
Alfresa NS-Prime 55 4.8±2.5 52.1 0 29
Alfresa NS-Plus C 12 56.5±6.2 11.0 10 66
Kyowa HM-Jack 32 0 0 0 7
SD Standard 27 25.0±0 0 0 31
CS-21-05 All 316 1.4±2.8 200.0 0 270
Eiken Oc sensor 183 0 0 0 270
Alfresa NS-Prime 55 5.8±3.0 51.7 0 16
Alfresa NS-Plus C 12 60.4±7.2 11.9 8 78
Kyowa HM-Jack 32 0 0 0 7
SD Standard 27 25.0±0 0 0 25
CS-21-06 All 316 292.8±22.1 7.5 11 1,000
Eiken Oc sensor 183 293.8±18.7 6.4 11 395
Alfresa NS-Prime 55 295.4±17.3 5.9 187 342
Alfresa NS-Plus C 12 251.9±13.7 5.4 130 267
Kyowa HM-Jack 32 78.2±17.5 22.4 35 101
SD Standard 27 586.0±99.2 16.9 416 732

Unusual results are indicated in bold. The instruments used were as follows: Eiken (Eiken Chemical Co., Tokyo, Japan), Kyowa (Kyowa Chemical Industry Co., Kagawa, Japan), Alfresa (Alfresa Pharma Co., Osaka, Japan), and SD (Standard Diagnostics, Seoul, Korea).

Abbreviations: SD, standard deviation; CV, coefficient of variation.

Lab Med Qual Assur 2023;45:1~9 https://doi.org/10.15263/jlmqa.2023.45.1.1