Laboratory Medicine and Quality Assurance

Table. 3.

Mean bias of participants of the accuracy-based creatinine proficiency testing program for creatinine

Year-round Specimen name No. of participants Target value of creatinine (mg/dL) Bias

Mean±SD (mg/dL) Mean±SD (%)
All data
2022-1st CAC-22-01 1,687 0.84 0.045±0.314 5.4±37.4
CAC-22-02 1,687 1.36 0.002±0.104 0.2±7.7
CAC-22-03 1,687 1.82 –0.003±0.149 –0.2±8.2
2022-2nd CAC-22-04 1,706 0.99 0.032±0.382 3.2±38.5
CAC-22-05 1,706 1.44 0.010±0.588 0.7±40.9
CAC-22-06 1,706 2.00 –0.013±0.127 –0.6±6.3
Without outliers*
2022-1st CAC-22-01 1,596 0.84 0.024±0.072 2.8±8.6
CAC-22-02 1,608 1.36 0.000±0.076 0.0±5.6
CAC-22-03 1,591 1.82 0.003±0.084 0.1±4.6
2022-2nd CAC-22-04 1,615 0.99 0.012±0.065 1.2±6.5
CAC-22-05 1,644 1.44 –0.004±0.074 –0.3±5.1
CAC-22-06 1,599 2.00 –0.007±0.087 –0.3±4.4

Abbreviations: SD, standard deviation; IQR, interquartile range; Q1, 25th percentile; Q3, 75th percentile.

*The outliers were defined as less than Q1–1.5×IQR or greater than Q3+1.5×IQR and were excluded from analysis.

Lab Med Qual Assur 2023;45:10~17 https://doi.org/10.15263/jlmqa.2023.45.1.10