Laboratory Medicine and Quality Assurance

Table. 2.

Mean bias of creatinine assay in the accuracy-based creatinine proficiency testing program (2023)

Year-round Sample name No. of participants
without outlier*
Target value of creatinine
(mg/dL)
Bias of creatinine assay
Mean±SD (mg/dL) Mean±SD (%)
2023-1st CAC-23-01 1,656 0.78 0.022±0.065 2.8±8.4
CAC-23-02 1,675 1.33 0.002±0.072 0.1±5.4
CAC-23-03 1,658 1.79 −0.008±0.079 −0.5±4.4
2023-2nd CAC-23-04 1,712 0.91 0.022±0.066 2.4±7.2
CAC-23-05 1,709 1.35 0.001±0.075 0.0±5.5
CAC-23-06 1,636 1.86 0.000±0.084 0.0±4.5

Abbreviations: SD, standard deviation; IQR, interquartile range; Q1, 25th percentile; Q3, 75th percentile.

*The outliers were defined as less than Q1−1.5×IQR or greater than Q3+1.5×IQR and were excluded from the analysis.

Lab Med Qual Assur 2024;46:78~86 https://doi.org/10.15263/jlmqa.2024.46.2.78