Laboratory Medicine and Quality Assurance

Table. 5.

Evaluation of concordance between reference products and the INCLIX TRF COVID-19 & Flu A/B Ag test

SARS-CoV-2 (SGTi-flex COVID-19 Ag) Influenza A (SGT Influenza A&B) Influenza B (SGT Influenza A&B)
INCLIX TRF COVID-19 & Flu A/B Ag
Overall concordance % (95% CI) 99.76 (98.63–99.96) 99.76 (98.63–99.96) 99.76 (98.63–99.96)
Positive percent agreement % (95% CI) 100.00 (95.36–100.00) 100.00 (95.72–100.00) 100.00 (95.99–100.00)
Negative percent agreement % (95% CI) 99.70 (98.30–99.95) 99.69 (98.27–99.95) 99.68 (98.23–99.94)
Kappa value 0.9922 0.9927 0.9930

Abbreviations: COVID-19, coronavirus disease 2019; SARS-CoV-2, severe acute respiratory syndrome coronavirus 2; CI, confidence interval.

Lab Med Qual Assur 2024;46:208~213 https://doi.org/10.15263/jlmqa.2024.46.4.208
© Lab Med Qual Assur