BK virus quantitative PCR external proficiency test results of the Korean Association of External Quality Assessment Service

Sample Reagent: Real-Q BK Virus Quantification Kit Reagent: all
No. (%)* Reported result range Median CV (%) No. (%)* Reported result range Median CV (%)
GV1-23-53 12 (41) 3.75–4.32 4.05 4.3 29 (100) 2.97–4.32 3.93 10.4
GV1-23-54 2.83–3.53 3.21 6.2 2.06–3.53 2.96 14.7
GV1-24-24 12 (46) 3.79–4.21 3.89 3.7 26 (100) 2.13–4.68 3.89 6.8
GV1-24-25 4.80–5.58 5.26 3.9 3.33–5.96 5.44 5.2

Abbreviations: PCR, polymerase chain reaction; CV, coefficient of variation.

*Number (percentage) of laboratories using specified reagent. The units for GV1-23-53 and GV1-23-54 are log10 copies/mL, while the units for GV1-24-24 and GV1-24-25 are log10 IU/mL.

Lab Med Qual Assur 2024;46:225~229 https://doi.org/10.15263/jlmqa.2024.46.4.225