Lab Med Qual Assur 2021; 43(4): 176-184

Published online December 31, 2021

https://doi.org/10.15263/jlmqa.2021.43.4.176

Copyright © Korean Association of External Quality Assessment Service.

Report of the Korean Association of External Quality Assessment Service on Cardiac Marker Testing (2016–2020)

Jong Do Seo1 , Hanah Kim1,2 , and Yeo-Min Yun1,2

1Department of Laboratory Medicine, Konkuk University Medical Center; 2Department of Laboratory Medicine, Konkuk University School of Medicine, Seoul, Korea

Correspondence to:Yeo-Min Yun
Department of Laboratory Medicine, Konkuk University Medical Center, Konkuk University School of Medicine, 120-1 Neungdong-ro, Gwangjin-gu, Seoul 05030, Korea
Tel +82-2-2030-5582
E-mail ymyun@kuh.ac.kr

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Cardiac marker tests are useful tools to diagnose acute coronary syndrome and predict its prognosis; therefore, reliable test results are required. The Korean Association of External Quality Assessment Service has conducted a proficiency test program for cardiac marker tests twice a year since 2016 using low- and medium-level pooled serum and high-level commercial quality control (QC) material. This report summarizes the results of the proficiency tests performed from 2016 to 2020. The number of participating laboratories increased from 321 in 2016 (1st trial) to 476 in 2020 (2nd trial). Items that showed the highest increase in participation during this period were tests for N-terminal-pro brain natriuretic peptide (NT-proBNP) (+86.2%) and creatine kinase-myocardial band (CK-MB) mass tests (+73.2%), whereas the corresponding BNP (–6.6%) and CK-MB activity (–26.5%) items were decreased. The most and least commonly tested items were CK-MB mass in 417 laboratories, and CK-MB activity in 25 laboratories, respectively. The coefficients of variation (CVs) differed among the manufacturers, and among different instruments from the same manufacturer. In 2020, during the 2nd trial, the lowest CV was confirmed in the troponin T test using the same manufacturer’s instruments, for the medium-level pooled serum sample. However, the CV for high-level commercial QC material was higher. Similarly, the NT-proBNP test for commercial QC material showed higher CV compared to pooled serum. These results are considered to originate from the non-commutability of commercial QC material; therefore, obtaining the commutability of the materials for the external quality assessment program is required.

Keywords: Cardiac marker, External quality assessment, Commutability

급성관상동맥증후군(acute coronary syndrome, ACS)은 빠른 진단이 예후에 큰 영향을 미치며, 감별 진단에는 심장표지자(cardiac marker)검사가 매우 중요하다. 미국심장협회의 진료지침에 따르면 ACS가 의심되는 환자에서 증상 발현 6시간 이내에 심장 트로포닌 검사를 시행하여 농도를 확인하고, 초기 검사에서 정상 농도를 나타낸 환자에서 심전도의 변화가 관찰되거나 ACS 관련 임상증상이 발생한 경우 증상 발현 6시간 이내에 추가 검사를 시행하여 ACS 진단에 활용할 것을 권고하고 있다. 또한 급성심근괴사 진단 시에는 트로포닌 최고농도뿐만 아니라 연속적인 변화에 대한 평가의 시행을 권고하는데, 초기 측정 농도가 참고치인 99백분위수 상한값보다 높으면서 추가 측정에서 연속으로 농도의 증가가 관찰되거나 20% 이상의 급격한 감소가 보이는 경우, 혹은 초기 측정 농도가 참고치보다는 낮거나 근접하다가 추가 측정에서 검사실 표준편차의 3배 이상 변동을 보이는 경우 급성심근괴사를 진단할 수 있다[1].

따라서 ACS 및 급성심근괴사를 진단할 때 심장표지자검사의 참고치에 해당하는 저농도에서의 검사 민감도와 정확도 확보가 중요하며, 외부정도관리 프로그램을 통하여 임상검사실의 심장표지자검사 능력을 평가하는 것은 심장질환 임상진료의 질 향상을 위하여 매우 중요하다.

대한임상검사정도관리협회는 2015년까지 운영해오던 분과 프로그램의 운영을 2016년도부터 차세대 신빙도조사사업 체계로 변경 출범하면서 심장표지자단백검사에 대한 신빙도조사를 임상화학 대분류의 단백검사 중분류 아래의 소분류 항목으로서 협회의 정규프로그램으로 집행하고 있다. 현재 심장표지자단백검사에는 9종의 검사항목이 포함되어 있다. 저자들은 본 보고서에서 2016년도부터 2020년까지 실시되었던 심장표지자단백검사의 신빙도조사 결과를 분석하여 보고하고자 한다.

1. 정도관리물질

심장표지자단백검사 프로그램에서는 상반기와 하반기로 나누어 연간 2회에 걸쳐 회차별, 항목별로 3개 농도의 정도관리물질을 이용하여 신빙도조사를 실시하고 있으며, 정도관리물질로는 저농도, 중간 농도의 자가제조물질과 고농도의 상용화 물질이 사용되고 있다.

심장표지자단백검사 프로그램 평가용의 저농도, 중간 농도 자가제조물질은 Clinical and Laboratory Standards Institute (CLSI) C37-A 지침을 기반으로 잔여 검체 혼합 혈청(pooled serum)에 고농도 심장표지자단백 잔여 혈청 검체를 첨가(spiking)하여 제조하였다[2]. 첨가 검체는 신빙도조사에 포함된 심장표지자단백 항목에 대한 임상검사가 의뢰되어 고농도의 측정결과가 확인된 환자의 잔여 혈청 검체를 별도로 수집하여 냉장보관하였던 것을 사용하였다. 혼합 혈청에 고농도 잔여 혈청 검체를 첨가한 후 4℃ 냉장실에서 교반기를 이용하여 16시간 이상 혼합하고, 혼합물을 3,000 rpm에서 10분간 원심분리 시행한 후 0.2 μm 필터를 통하여 잔여 침전물을 제거하여 제조한 저농도와 중간 농도의 외부정도관리물질을 농도별로 각각 1 mL 바이알에 분주하여 냉동보관하였다.

고농도 외부정도관리물질은 상용화 물질인 Seronorm Cardiac Acute Liq (SERO AS Stasjonsveien 44, NO-1396; SERO AS, Billingstad, Norway)를 이용하였으며, 신빙도조사사업 각 회차마다 상술한 2개 농도의 자가제조물질과 상용화 물질 1개로 평가물질 세트를 구성하여 심장표지자단백검사 프로그램 참가기관에 발송하였다.

2. 신빙도조사 심장표지자 종목

2016년도부터 2020년도까지 시행한 대한임상검사정도관리협회의 심장표지자단백검사 신빙도조사에 포함된 심장표지자 종목은 troponin I (TnI), troponin T (TnT), creatine kinase-MB (CK-MB) activity, CK-MB mass, myoglobin, brain natriuretic peptide (BNP), N-terminal pro-brain natriuretic peptide (NT-proBNP), high-sensitivity C-reactive protein (hs-CRP), homocysteine으로 5년간 1회차와 2회자 모두 동일한 종목을 대상으로 신빙도조사를 시행하였다.

3. 결과분석 및 판정

신빙도조사 평가물질은 택배를 이용하여 참가기관에 발송하였고, 각 기관에서 입력한 결과는 전체 참여기관, 장비 제조회사별, 측정 기기별로 입력된 결과에 대하여 통계분석을 시행하였고 참가기관에는 개별기관의 평가결과를 정리한 기관별 보고서와 프로그램 참여기관 전체의 결과를 정리한 공통 보고서가 제공되었다. 공통 보고서에는 평가종목별로 전체 참여기관(all), 장비 제조사에 기초한 기준분류, 측정기기에 기초한 세분류에 따라 참여기관의 수(N), 측정값의 평균(mean), 측정값의 표준편차(standard deviation), 변이계수(coefficient of variation, CV), 중간값(median), 최소값(min), 최대값(max)을 제시하였다. 범용 장비와 범용 시약을 사용하는 검사시스템의 기준분류는 2019년도 1차 신빙도조사까지는 장비 제조사를 기준으로 분류하였으나 2019년도 2차 신빙도조사부터는 시약 제조사를 기준으로 하는 분류로 변경되었다.

평균, 표준편차, 변이계수는 분류별 분석에서 75백분위수의 값(Q3)과 25백분위수의 값(Q1) 사이의 차이값(Q3–Q1; 사분위수 범위, interquartile range, IQR)의 1.5배를 초과하여 Q1보다 작거나 Q3보다 큰 결과값(<Q1–1.5×IQR or >Q3+1.5×IQR)을 이상치로 간주하여 제거한 후 분석한 값을 제시하였고 중간값, 최소값, 최대값은 이상치를 제거하지 않고 해당 분류에 속한 기관을 모두 포함하여 제시하였다.

분류에 속하는 기관 수가 10개 이상이면서 이상치 제거 후 남은 값이 8개 이상인 경우에는 상술한 지표를 모두 보고서에 제시하였고, 참여기관이 10개 미만이거나 이상치 제거 후 남은 값이 8개 미만인 경우는 평균, 표준편차, 변이계수를 제시하지 않고 중간값, 최소값, 최대값만을 표기하였다. 참여기관이 2개인 경우는 중간값을 제시하지 않고 낮은 측정값과 높은 측정값을 각각 최소값과 최대값으로 제시하였으며, 기관 수가 1개인 경우는 2017년도까지의 공통 보고서에는 포함되지 않았으나 2018년도부터의 공통 보고서에는 포함되어 입력된 값을 중간값으로 보고하였다.

기관별 보고서에는 각 기관의 입력결과를 바탕으로 기준분류 및 세분류에 대하여, 상술한 평균 및 표준편차 제시조건을 충족하는 경우 표준편차지수(standard deviation index, SDI)를 함께 제시하였고, 기준분류 및 세분류의 SDI값이 ±3 이내인 경우를 적합(acceptable)한 것으로 판정하였다.

1. 신빙도조사 참여기관

2016년도 1회차부터 2020년도 2회차까지 심장표지자단백 신빙도조사사업에 참여하여 결과를 회신한 기관의 수는 321개 기관에서 476개 기관으로 증가하였으며, 회신율은 97.5%–99.8%였다. 2016년도 1회차 심장표지자단백검사 프로그램은 324개 기관에 검체를 발송하여 321개(99.1%) 기관에서 회신하였으며 2020년도 1회차 프로그램은 473개 기관에 검체를 발송하여 468개 기관(98.9%)에서 회신하였고, 2회차에 477개 기관에 검체를 발송하여 476개 기관(99.8%)에서 회신하였다.

Table 1에 각 검사항목별 신빙도조사 참여기관과 함께 10개 이상의 기관이 포함된 주요 기준분류의 수를 표시하였으며, 2016년도 1회차와 비교하여 2020년도 2회차 참여기관 수의 변화를 살펴보았을 때 NT-proBNP (+86.2%), CK-MB mass (+73.2%) 항목에 대한 참여기관의 증가가 두드러졌으며, 각각에 대응하는 BNP (–6.6%), CK-MB activity (–26.5%) 항목은 참여기관 수에 감소가 있었다. 나머지 5개 평가항목에 대해서는 +4.5%에서 +44.3%의 참여기관 증가가 확인되었다. 2020년도 2회차의 심장표지자단백검사 프로그램에서 가장 많은 기관이 참여한 심장표지자 종목은 CK-MB mass로 417개 기관에서 참여하였으며 가장 적은 기관이 참여한 종목은 25개 기관에서 참여한 CK-MB activity였다.

Table 1 . Number of laboratories that participated in each cardiac marker proficiency test from 2016 to 2020.

Test substanceCriteria classificationYear% Change (2020-2nd vs. 2016-1st)

20162017201820192020





1st2nd1st2nd1st2nd1st2nd1st2nd
TnIAll25925628931334334235836035436239.8
Siemens62626061595656524949
Abbott59566867767978767683
Roche40404854606368727070
Beckman Coulter33333635363435363638
LSI Medience Corp.22242831343334343433
Radiometer19192934424548525152
Biomerieux131313191815151297
Boditech Med Inc.23513141618
TnTAll7979899810810511010811111444.3
Roche78798998107104109108111114
CK-MB (activity)All34343735363529252825–26.5
Roche111213101195443
Beckman Coulter7767669978
Hitachi*7791112139
Open system*6108
CK-MB (mass)All27227330734037337640440940741773.2
Roche8587101116128134146153156157
Siemens60595961605755514949
Abbott38375051596160605662
Beckman Coulter33333433343237383840
LSI Medience Corp.19212325282728282928
Radiometer18192934414347494849
Biomerieux99711888775
Boditech Med Inc.234991014
MyoglobinAll1341321391361381381441411351404.5
Roche43435052545656575454
Siemens33333130262628242523
Beckman Coulter18181817181717181617
Abbott17151816161615151521
Radiometer7776101011111010
BNPAll918891911059795878785–6.6
Abbott28253031343637343131
Siemens28272727312826232523
Beckman Coulter21232321191715151515
Alere Inc.14121110121213141615
NT-proBNPAll15915919020522423525826928129686.2
Roche9496113121128140145152156161
Biomerieux19151516161821181611
LSI Medience Corp.17212729323132302928
Radiometer15162327333441424245
Siemens88888810121519
Boditech Med Inc.1246810
Abbott1261119
hs-CRPAll13612913414114114315115015116017.6
Beckman Coulter32313541424446454545
Roche31303032313132363845
Siemens1098681013121211
Hitachi*30313433333336
Toshiba*24242323232321
Open system*394450
HomocysteineAll14014015115417217418017117317222.9
Abbott40373838414139393738
Roche25263132363943454742
Siemens22232323252522232123
Beckman Coulter21232828302833323231
Hitachi*22222123283032
Toshiba*6766766
Open system*182530

The manufacturers supported by 10 or more institutions at least once in the EQA evaluation period (2016–2020) are presented in the table. The percentage change in each analyte was calculated by comparing the total number of participants in the recent (2020-2nd) EQA program with that in the initial (2016-1st) EQA program. The instruments used were from the following companies: Siemens (Tarrytown, NY, USA), Abbott (Abbott Park, IL, USA), Roche (Mannheim, Germany), Beckman Coulter (Brea, CA, USA), LSI Medience Corp. (Tokyo, Japan), Radiometer (Brønshøj, Denmark), Biomerieux (Marcy l’Etoile, France), Boditech Med Inc. (Chuncheon, Korea), Hitachi (Chiyoda, Japan), Alere Inc. (Waltham, MA, USA), and Toshiba (Tokyo, Japan)..

Abbreviations: TnI, troponin I; TnT, troponin T; CK-MB, creatine kinase-MB; BNP, brain natriuretic peptide; NT-proBNP, N-terminal pro-brain natriuretic peptide; hs-CRP, high sensitivity C-reactive protein; EQA, external quality assessment..

*The institutions using open system chemistry analyzers were grouped by instrument manufacturers until the 2019-1st EQA program, and were grouped by reagent manufacturers after the 2019-2nd EQA program..



2. 신빙도조사 결과보고에 사용된 장비

심장표지자단백검사 신빙도조사 결과의 보고에는 대부분의 참가기관에서 면역측정장비를 사용하였으며, CK-MB activity, hs-CRP, homocysteine 등의 검사종목의 경우 일부 참가기관에서 일반화학측정장비를 사용하여 결과를 보고하였다. Table 1에 표시된 바와 같이 장비 제조사별 기준분류에서는 Roche (Mannheim, Germany) 검사장비를 사용하는 참가기관이 가장 많았고, Abbott (Abbott Park, IL, USA), Siemens (Tarrytown, NY, USA), Beckman Coulter (Brea, CA, USA), Radiometer (Brønshøj, Denmark) 검사장비도 다수의 참가기관에서 사용하고 있었다. 2020년도 2회차의 신빙도조사사업을 기준으로, 총 9개의 심장표지자단백검사 항목 중 TnT, CK-MB mass, myoglobin, NT-proBNP, homocysteine 등 5개 검사항목에 대해서는 Roche 장비를 사용하는 기관이 가장 많았고, TnI, BNP 등 2개 검사항목에서는 Abbott 장비, CK-MB activity, hs-CRP 등 2개 검사에서는 범용 장비를 사용하는 기관이 가장 많았다.

신빙도조사사업 초기부터 동일한 범용 장비와 범용 시약을 사용하여 지속적으로 참여한 기관의 경우 상술한 2019년도 2회차의 기준 변경에 따라 기준분류 및 세분류의 변화가 있을 수 있다.

3. 신빙도조사 결과의 통계

2020년도 2회차의 심장표지자단백검사의 검사항목별, 정도관리물질 농도별 분석결과를 평균과 표준편차, 변이계수의 산정이 가능하였던 주요 세분류를 대상으로 Table 2와 같이 정리하였다. 전체 참가기관에 대하여 가장 작은 검사실 간 변이계수를 보인 검사항목은 TnT 검사항목의 CCA-20-05 물질이었으며(3.6%), 이는 해당 검사항목의 경우 모든 참가기관이 동일 제조사(Roche)의 검사장비를 사용하기 때문인 것으로 생각되었다. 반면에 다양한 제조사의 장비를 사용하는 기관이 참가한 TnI 검사항목의 경우 3개 농도의 평가물질 세트에 대하여 전체 참가기관 간에 51.6%–77.1%의 큰 변이계수를 보였다.

Table 2 . Results of cardiac marker proficiency test for subgroups during the 2020-2nd EQA program.

Test substanceInstrumentNo. of labsCCA-20-04 (pooled human serum)CCA-20-05 (pooled human serum)CCA-20-06 (commercial control material)



Mean±SDCV (%)Mean±SDCV (%)Mean±SDCV (%)
TnIAll3620.179±0.13877.10.227±0.17476.85.220±2.69651.6*
ADVIA Centaur XPT120.362±0.0298.10.468±0.0388.017.602±0.6713.8
ARCHITECT i1000220.360±0.0267.10.452±0.0337.37.359±0.5497.5
ARCHITECT i2000560.365±0.0246.70.452±0.036.77.142±0.4356.1
UniCel DxI800340.022±0.00210.00.026±0.0029.25.253±0.1963.7
Pathfast330.051±0.00714.40.068±0.00913.94.839±0.50710.5
AFIAS-6170.175±0.0158.30.220±0.0188.29.704±0.8358.6
AQT90 FLEX520.022±0.0028.20.026±0.00310.81.153±0.0575.0
cobas4000 e411410.164±0.03219.50.207±0.03215.53.945±0.3328.4
cobas6000 e601170.159±0.02113.20.200±0.02412.23.596±0.1754.9
TnTAll1140.156±0.0063.70.197±0.0073.61.674±0.53131.7*
cobas e801290.157±0.0053.00.199±0.0052.52.116±0.1627.7
cobas4000 e411510.156±0.0064.00.199±0.0094.41.124±0.12911.5
cobas6000 e601150.154±0.0053.30.196±0.0073.32.05±0.20910.2
cobas8000 e602190.153±0.0063.70.194±0.0052.62.028±0.29.9
BNPAll85NANANANA1,382.0±167.312.1
ARCHITECT i200023NANANANA1,414.4±143.610.2
UniCel DxI80015NANANANA1,265.5±41.53.3
Triage Meter-pro15NANANANA615.4±109.517.8
NT-proBNPAll2966,295.9±360.35.78,020.3±412.75.24,604.6±2,399.052.1*
cobas e801386,042.1±162.92.77,795.5±199.22.63,287.5±122.73.7
cobas4000 e411686,282.7±228.13.68,134.6±330.74.12,705.3±161.56.0
cobas6000 e601296,242.5±221.63.67,922.2±225.42.83,415.1±236.66.9
cobas8000 e602266,158.7±183.83.07,958.3±238.33.03,425.1±240.47.0
Pathfast285,820.1±657.011.37,588.7±712.59.410,402.2±8087.8
AQT90 FLEX456,462.3±250.23.98,115.8±365.44.56,767.2±291.24.3
AFIAS-6106,623.4±540.98.28,660.5±678.17.820,055.5±1,957.29.8
ARCHITECT i2000167,335.1±336.94.69,370.6±464.85.04,120.6±232.25.6

The instruments used in 10 or more institutions during the 2020-2nd EQA program are presented in the table. The level of BNP, which in ethylenediaminetetraacetic acid-plasma is a suitable specimen recommended by the test manufacturer, was measured in commercial material only. The instruments used were from the following companies: Siemens (Tarrytown, NY, USA), Abbott (Abbott Park, IL, USA), Beckman Coulter (Brea, CA, USA), Polymedco (Cortlandt Manor, NY, USA), Boditech Med Inc. (Chuncheon, Korea), Radiometer (Brønshøj, Denmark), Roche (Mannheim, Germany), and Quidel Corp. (San Diego, CA, USA)..

Abbreviations: EQA, external quality assessment; SD, standard deviation; CV, coefficient of variation; TnI, troponin I; TnT, troponin T; BNP, brain natriuretic peptide; NA, not available; NT-proBNP, N-terminal pro-brain natriuretic peptide..

*CCA-20-06, the commercial control material with high concentration of cardiac marker, showed higher CV% compared to other materials..



2016년도 1회차부터 2020년도 2회자까지 10회에 걸친 신빙도조사사업에서 중간 농도 자체 제조 평가물질과 고농도 상용화 평가물질에 대한 검사항목 TnI, TnT, BNP, NT-proBNP의 주요 기준분류 별 변이계수 추이는 Fig. 1에 나타낸 것과 같다. 일부 기준분류에서 Q1과 Q3이 동일한 특정 값인 경우, 기준에 따라 해당 값 이외의 입력결과는 이상치로 분류되어 제거되었으며 표준편차와 변이계수는 0으로 계산되었다.

Figure 1. Change in coefficient of variation % during the period of cardiac marker external quality assessment program from 2016-1st to 2020-2nd for (A, B) troponin I (TnI) in pooled serum and commercial control material, respectively. (C, D) Troponin T (TnT), (E, F) brain natriuretic peptide (BNP), and (G, H) N-terminal pro-brain natriuretic peptide (NT-proBNP) of each manufacturer. BNP test for the pooled serum sample has been discontinued since 2018 because ethylenediaminetetraacetic acid-plasma is the only suitable specimen recommended by the manufacturers. The instruments used were from the following companies: Siemens (Tarrytown, NY, USA), Abbott (Abbott Park, IL, USA), Beckman Coulter (Brea, CA, USA), LSI Medience Corp. (Tokyo, Japan), Biomerieux (Marcy l’Etoile, France), Radiometer (Brønshøj, Denmark), Roche (Mannheim, Germany), and Alere Inc. (Waltham, MA, USA) (Continued on next page).

2020년도 2회차의 심장표지자단백검사 신빙도조사사업에서 TnT 검사항목의 CCA-20-05 물질에서 3.6%로 가장 작은 변이계수를 나타냈고, 이는 모든 참가기관이 동일 제조사의 장비를 사용하기 때문인 것으로 생각되었다. 단, 고농도 표준물질인 CCA-20-06의 검사에서는 동일 제조사의 장비를 사용함에도 불구하고 31.7%의 높은 변이계수를 나타냈는데, 이는 세분류상 cobas 4000 장비에서 다른 장비에 비하여 유의하게 낮은 측정값을 나타낸 것이 원인인 것으로 생각되었다. NT-proBNP의 측정에서도 고농도 표준물질에서 저농도 및 중간 농도 표준물질에 비하여 높은 변이계수를 보이는 양상이 확인되었으며(52.1% vs. 5.2%–5.7%), 해당 검사항목의 경우 기준분류 내의 측정값 차이는 크지 않은데 비하여 장비 제조사 간 차이가 크게 나타났다. 이는 CLSI 지침에 따라 교환 가능성(commutability)의 확보가 가능하도록 자체 제조한 저농도 및 중간 농도의 평가물질에 비하여 상용화 물질인 고농도 평가물질의 경우 검사시스템 간 교환 가능성을 확보하지 못하여 나타난 현상으로 추정되며, 외부정도관리를 시행할 때 평가물질의 교환 가능성 확인에 유의할 필요가 있을 것으로 생각된다.

3개 농도의 평가물질 세트에 대하여 전체 참가기관 간에 큰 변이계수를 나타낸 TnI 검사의 경우, 동일한 검사기기를 사용하는 세분류 그룹의 변이계수는 cobas 4000 장비를 사용한 CCA-20-04의 측정에서 최대 19.5%로 전체 기관간 변이계수에 비하여 작아, 검사기기 간의 측정값 차이가 높은 변이계수에 크게 기여한 것으로 판단되었다. 이는 참여기관의 검사 수행능의 차이에서 기반한 영향보다는 각 검사실에서 사용하고 있는 검사의 면역측정 항체의 특성이나 평가물질 기질에 대한 각 제조사 시약의 반응성 차이(matrix effect)에 의한 결과로 추정된다.

이에 따라 향후의 심장표지자단백검사 신빙도조사사업에서는 각 제조사의 검사방법 간의 기질효과를 최소화하여 임상검사실 심장표지자검사의 신뢰도를 보다 정확하게 평가할 수 있도록, 평가에 활용되는 자체 제조 표준물질과 상용화 표준물질을 대상으로 한 교환 가능성의 사전 평가를 검토할 필요가 있을 것으로 생각된다[3].

  1. Amsterdam EA, Wenger NK, Brindis RG, Casey DE Jr, Ganiats TG, Holmes DR Jr, et al. 2014 AHA/ACC guideline for the management of patients with non-ST-elevation acute coronary syndromes: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. Circulation 2014;130:2354-94.
    Pubmed CrossRef
  2. Danilenko U, Vesper HW, Myers GL, Clapshaw PA, Camara JE, Miller WG. An updated protocol based on CLSI document C37 for preparation of off-the-clot serum from individual units for use alone or to prepare commutable pooled serum reference materials. Clin Chem Lab Med 2020;58:368-74.
    Pubmed KoreaMed CrossRef
  3. Clinical and Laboratory Standards Institute. Evaluation of commutability of processed samples; approved guideline: EP14-A3. 3rd ed. Wayne (PA): Clinical and Laboratory Standards Institute; 2014.