• Sitemap
  • Contact Us

pISSN 2950-9114 eISSN 2950-9122
Article View

Annual Report

J Lab Med Qual Assur 2015; 37(4): 190-208

Published online December 31, 2015

https://doi.org/10.15263/jlmqa.2015.37.4.190

Copyright © Korean Association of External Quality Assessment Service.

Annual Report on External Quality Assessment Scheme for Immunoassay Tests in Korea (2014)

Hyon-Suk Kim and Young Lan Kim, as Immunoassay Subcommittee, Korean Association of External Quality Assessment Service

Department of Laboratory Medicine, Yonsei University College of Medicine, Seoul, Korea

Correspondence to:Hyon-Suk Kim
Department of Laboratory Medicine, Yonsei University College of Medicine, 50-1 Yonseiro, Seodaemun-gu, Seoul 03722, Korea Tel: +82-2-2228-2443 Fax: +82-2-364-1583 E-mail: kimhs54@yuhs.ac

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

In 2014, two external quality assessment trials were performed on 13 test items grouped in four categories. The laboratories procured the materials for the first and second trials on 13 May 2014 and 11 November 2014, respectively. The trials were performed on 13 test items, including tumour markers, thyroid hormones, cardiac marker troponin (troponin T or troponin I), and procalcitonin as a new biomarker for immunoassay methods. The bone marker carboxy-terminal collagen crosslinks (CTX) has been replaced by procalcitonin this year because only a limited number of institutions used it. External quality surveys of the 13 immunoassay test items with 8 control materials were performed as scheduled. The 13 control materials included six tumour markers, alpha-fetoprotein, carcinoembryonic antigen, carcinoma antigen 125, carbohydrate antigen 19-9, human chorionic gonadotrophin, and prostate specific antigen, as well as five thyroid markers, thyroid hormone 3 (T3), T4, thyroid stimulating hormone, free T4, and thyroglobulin. This year, procalcitonin has been introduced as a new biomarker in addition to troponin, which was introduced last year. Five homemade pooled sera and three commercial control sera were used as survey materials. The MAS Tri-point Liquimmune level 3 (Medical Analysis Systems Inc., USA) was used for controls of thyroid hormones, while Elecsys PreciControl Varia and Elecsys PreciControl Troponin (Roche, Germany) were used for controls of the new biomarkers procalcitonin and troponin, respectively. In the external quality assessment by the Immunoassay Subcommittee, 712 institutions participated in the first trial survey (response rate 97.9%), while 715 participated in the second survey (response rate 97.9%). The quality of the participating laboratories seems to be continuously improving compared to the results of their peers. Additionally, this year procalcitonin has been introduced as a new biomarker instead of the CTX, which was used in 2013, while thyroglobulin and troponin-T/troponin-I, which were used for the 2013 samples, continue to be used in surveys by the Immunoassay Subcommittee.
(J Lab Med Qual Assur 2015;37:190-208)

Keywords: Immunoassay, Quality assurance, External quality assessment, Survey, Tumour marker, Thyroid marker, Cardiac marker, Procalcitonin

  1. Kim HS, Kim YL; Immunoassay Subcommittee, The Korean Association of Quality Assurance for Clinical Laboratories. Annual report on external quality assessment of immunoassay testing in Korea (2013). J Lab Med Qual Assur 2014;36:171-89.
    CrossRef
  2. Kim HS, Kim YL, Park J, Kwon OH, Kim DA, Kim JQ, et al. Immunoassay external quality assessment of immunoassay subcommittee (2010). J Lab Med Qual Assur 2011;33(Suppl):S91-114.
  3. Kim HS, Kim YL, Park J, Kwon OH, Kim DA, Kim JQ, et al. Annual report on external quality assessment in immunoassay in Korea (2009). J Lab Med Qual Assur 2010;32:103-14.
  4. Cha YJ, Kwon SY, Kim TY, Kim JR, Kim HS, Park MH, et al. Annual report on external quality assessment in immunoserology in Korea (2008). J Lab Med Qual Assur 2009;31:49-72.
  5. Kim HS, Kim YL, Lim HS, Park J, Chun HS, Shin S, et al. Annual report on external quality assessment of Immunoassay Subcommittee in Korean Clinical Laboratory Survey (2007). J Lab Med Qual Assur 2008;30:111-32.
  6. Kim HS, Kim YL, Lim HS, Chun HS, Shin S, Jung YS, et al. Annual report on external quality assessment of Immunoassay Subcommittee in Korean Clinical Laboratory Survey (2006). J Lab Med Qual Assur 2007;29:99-119.
  7. Kim HS. Immunoassay external quality control of immunoassays. J Lab Med Qual Assur 2003;25(S2): S38791.
  8. Kim HS, Lim HS, Kwon OH, Kim DA, Kim YK, Kim YR, et al. Annual report on external quality assessment of immunoassay in Korea (2001). J Clin Pathol Qual Control 2002;24:83-94.
  9. Kwon OH, Lim HS, Kim HS, Kim DA, Kim YK, Kim JQ, et al. Annual report on external quality assessment in immunoassay in Korea (2000). J Clin Pathol Qual Control 2001;23:111-44.
  10. Kim HS. Future trends of immunoassay EQA. J Clin Pathol Qual Control 2001;23:S275-6.
  11. Bock JL. The new era of automated immunoassay. Am J Clin Pathol 2000;113:628-46.
    Pubmed CrossRef
  12. Ritchie RF. Immunoanalytical standardization: a global science. Scand J Clin Lab Invest Suppl 1991;205:3-10.
    Pubmed CrossRef
  13. Whicher JT. Calibration is the key to immunoassay but the ideal calibrator is unattainable. Scand J Clin Lab Invest Suppl 1991;205:21-32.
    Pubmed CrossRef
  14. Wood WG. “Matrix effects” in immunoassays. Scand J Clin Lab Invest Suppl 1991;205:105-12.
    Pubmed CrossRef
  15. Tahlan A, Barrowcliff e T, Das RG, Shin J, Wood D. WHO technical workshop on stability of reference materials for biological medicines and in vitro diagnostics, Geneva, Switzerland, 28-29 November 2005. Biologicals 2007;35:361-5.
    Pubmed CrossRef
  16. Jenny RW, Jackson-Tarentino KY. Causes of unsatisfactory performance in proficiency testing. Clin Chem 2000;46:89-99.
    Pubmed
  17. Kwon OH, Lim HS, Kim HS, Kim DU, Kim YK, Kim JQ, et al. Annual report on external quality assessment in immunoassay in Korea (1995). J Clin Pathol Qual Control 1996;18:95-117.
  18. Kim HS. Internal quality control of immunochemistry tests. J Clin Pathol Qual Control 1993;15:S147-51.
  19. Feldkamp CS, Carey JL. Standardization of immunoassay methodologies. In: Rose NR, Hamilton RG, Detrick B, editors. Manual of clinical laboratory immunology. 6th ed. Washington (DC): ASM Press, 2002: 1215-26.
  20. College of American Pathologists. Standards for laboratory accreditation: Laboratory Accreditation Program 1998 edition. http://www.cap.org (AccessedDecember 2, 2015).
  21. Miyai K, Price CP. Problems for improving performance in immunoassay. J Int Fed Clin Chem 1992;4:154-63.
    Pubmed

Supplementary File


Share this article on :

Stats or metrics