• Sitemap
  • Contact Us

pISSN 2950-9114 eISSN 2950-9122
Article View

Original Article

J Lab Med Qual Assur 2016; 38(4): 225-233

Published online December 31, 2016

https://doi.org/10.15263/jlmqa.2016.38.4.225

Copyright © Korean Association of External Quality Assessment Service.

Evaluation of the i-STAT Point-of-Care Analyser

YeJin Oh, Sun Hoe Koo, and Gye Cheol Kwon

Department of Laboratory Medicine, Chungnam National University Hospital, Chungnam National University School of Medicine, Daejeon, Korea

Correspondence to:Gye Cheol Kwon
Department of Laboratory Medicine, Chungnam National University Hospital, Chungnam National University School of Medicine, 282 Munhwa-ro, Junggu, Daejeon 35015, Korea
Tel: +82-42-280-7799 Fax: +82-42-257-5365 E-mail: kckwon@cnu.ac.kr

Received: April 12, 2016; Revised: August 10, 2016; Accepted: August 23, 2016

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Background: Point-of-care testing (POCT) is designed to be used near the site where the clinical care is being delivered. The demand for POCT in the medical field is expanding significantly, given that rapid results can eventually lead to early diagnosis and immediate clinical management of diseases. Therefore, the aim of this study was to evaluate the performance of the i-STAT POC analyser (Abbott Diagnostics, USA) for testing 8 chemical analytes (viz., sodium, potassium, chloride, total carbon dioxide, blood urea nitrogen, creatinine, glucose, and ionised calcium) and 2 hematological analytes (hematocrit [HCT], hemoglobin [Hb]).
Methods: The precision and linearity of the 10 analytes were measured according to Clinical and Laboratory Standards Institute (CLSI) EP15-A3 and EP6-A guidelines. Comparisons with a central laboratory hematology analyser, Coulter LH 780 (Beckman Coulter Inc., USA), and a chemical analyser, UniCel DxC 880i (Beckman Coulter Inc.), were performed using 85 patient samples according to CLSI EP9-A3.
Results: The coefficient of variation values for the within-run precision and total precision at 3 levels of all analytes were within 5%, except those for low level creatinine. In the aspect of linearity, the correlation coefficient values of all analytes were over 0.975 in the clinically important concentration range. A very high correlation was observed in glucose, blood urea nitrogen and creatinine (R>0.975), high correlation was observed in sodium, potassium, Hct and Hb (R>0.9), and relatively good correlation was observed in chloride and total carbon dioxide (R>0.7) compared to the central laboratory analysers.
Conclusions: i-STAT showed relatively high precision and linearity, and comparable data to that of routine hematology and chemistry analysers. This device was concluded to have potential for providing faster results and relatively acceptable values to clinicians in need of immediate results.

Keywords: Point-of-care systems, i-STAT, Evaluation studies

  1. Clinical and Laboratory Standards Institute. Selection criteria for point-of care testing devices: approved guideline. Wayne (PA): Clinical and Laboratory Standards Institute, 2010.
  2. Jeong TD, Kang HJ, Kim MS, Kim SY, Lee W, Chun S, et al. Evaluation of the performance of LABGEO PT Hepatic Test 9. Lab Med Online 2013;3:138-44.
    CrossRef
  3. Park H, Ko DH, Kim JQ, Song SH. Performance evaluation of the Piccolo xpress Point-of-care Chemistry Analyzer. Korean J Lab Med 2009;29:430-8.
    Pubmed CrossRef
  4. Dascombe BJ, Reaburn PR, Sirotic AC, Coutts AJ. The reliability of the i-STAT clinical portable analyser. J Sci Med Sport 2007;10:135-40.
    Pubmed CrossRef
  5. Yi SG, Lee SY, Kim JW. Evaluation of the i-STAT pointofcare testing analyzer. Korean J Lab Med 2002;22:30411.
  6. Nichols JH. Point of care testing. Clin Lab Med 2007;27:893-908.
    Pubmed CrossRef
  7. Clinical and Laboratory Standards Institute. User verification of performance for precision and trueness:approved guideline. 2nd ed. Wayne (PA): Clinical and Laboratory Standards Institute, 2005.
  8. Clinical and Laboratory Standards Institute. Evaluation of the linearity of quantitative measurement procedures:a statistical approach: approved guideline. Wayne (PA):Clinical and Laboratory Standards Institute, 2003.
  9. Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples:approved guideline. 2nd ed. Wayne (PA): Clinical and Laboratory Standards Institute, 2002.
  10. Statland BE. Clinical decision levels for laboratory tests. 2nd ed. Oradell (NJ): Medical Economics Books, 1987.
  11. Centers for Disease Control and Prevention. Clinical Laboratory Improvement Amendments regulation:standards and certification: laboratory requirements. http://wwwn.cdc.gov/clia/Regulatory/default.aspx(Accessed March 21, 2016).
  12. Kim DK, Shin H, Kim B, Jeong SH, Lim JB. Evaluation of i-STAT CHEM8+ Point-of-care Chemistry Analyzer. Lab Med Online 2015;5:57-62.
    CrossRef
  13. Zady RM. Z-STATS 12: correlation and simple least squares regression. https://www.westgard.com/lesson42.htm (Assessed March 21, 2016).

Supplementary File


Share this article on :

Stats or metrics